跳到主要导航 跳到搜索 跳到主要内容

Fraxicon composite fiber probe design for OCT imaging concurrent enhancement

  • University of Science and Technology Beijing
  • National Key Laboratory on Near-Surface Detection
  • Beijing Institute of Technology
  • Shenzhen Urban Public Safety and Technology Institute
  • China General Nuclear Power Group
  • State Key Laboratory of Nuclear Power Safety Monitoring Technology and Equipment

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

Optical coherence tomography (OCT) fiber probes are critical for assessing internal cavity organs and investigating pathogenesis, but balancing large starting focal length, extended depth of focus (DOF), and high lateral resolution remains challenging in miniaturized designs. Here, we present an ultra-compact (outer diameter ∼ 383 μm) optical probe with a hybrid micro-lens architecture to address these conflicts. Fabricated via two-photon polymerization (TPP) based on femtosecond laser direct writing, the design integrates a negative axicon and Fresnel axicon (fraxicon) at the tip of a fiber composite assembly, which employs a large diameter no core fiber (LD-NCF) to expand the beam from a single-mode fiber (SMF). The negative axicon enhances the starting focal length, while the fraxicon resolves the DOF-lateral resolution trade-off. The prototype achieves a ∼ 415 μm starting focal length, sub-5.71 μm lateral resolution over a 304 μm DOF, and validated performance in multi-sample OCT imaging, highlighting its commercial potential for internal cavity imaging.

源语言英语
文章编号114402
期刊Optics and Laser Technology
卷194
DOI
出版状态已出版 - 2月 2026

学术指纹

探究 'Fraxicon composite fiber probe design for OCT imaging concurrent enhancement' 的科研主题。它们共同构成独一无二的学术指纹。

引用此

Baidu
map